Bacteriological Incubator
Product Details
Product Description
“TEMPO” make bacteriological incubator are specially designed for research laboratories, engineering lab, pharmaceutical lab, cosmetic industry, chemical lab, clinical lab, microbiological determination, pharmaceutical stability assays, food processing, QC, etc.
- Temperature range – 5°C above ambient to 80°C
- Temperature accuracy – ±0.2°C
- Supply – 230V, AC
Construction
“TEMPO” make bacteriological incubator are made up of double walled construction. Full view inner acrylic door & front door with gasket & lock. The motorized blower is fitted on the backside for uniform temperature (only forced air circulation model). Insulation is provided by means of suitable PUF insulation. The unit is provided with adjustable stainless steel rod trays.
Technical Specification
Controlling System
Microprocessor based autotune PID temperature controller with a dual display of the value and process value for precise control of temperature with PT 100 sensor is used.
Air Circulation (only forced air circulation)
The motorized blower at the backside of the chamber develops a unique airflow system which ensures maximum uniform temperature distribution inside the chamber. A unique airflow ensures quick recovery after the doors are opened.
Heating System
Forced air circulation model – high quality SS tubular heaters are used for better heating conditions.
Natural convection model – the heating elements are made from the best quality Kanthal/Nichrome wire for refractory purpose.
Features
Full view observation acrylic door with a gasket to observe the sample inside the chamber. A specially designed stainless steel rod trays ensure uniform temperature distribution.
Safety Features
Safety standby thermostat to cut off heater supply in case of overshoot of temperature.
Models & Optional Features
Natural Convention Model
Models | Capacity (litres) | Internal Dimensions H × W × D (cm) | Trays |
---|---|---|---|
TI – 90 A | 43 | 35 × 35 × 35 | 2 |
TI – 90 B | 91 | 45 × 45 × 45 | 2 |
Forced Air Circulation Model
Models | Capacity (litres) | Internal Dimensions H × W × D (cm) | Trays |
---|---|---|---|
TI – 90 A S/G | 43 | 35 × 35 × 35 | 2 |
TI – 90 B S/G | 91 | 45 × 45 × 45 | 2 |
TI – 90 C S/G | 100 | 50 × 50 × 40 | 2 |
TI – 90 D S/G | 170 | 70 × 60 × 40 | 2 |
TI – 90 E S/G | 200 | 60 × 60 × 60 | 2 |
TI – 90 F S/G | 285 | 80 × 60 × 60 | 3 |
TI – 90 G S/G | 325 | 90 × 60 × 60 | 4 |
TI – 90 H S/G | 400 | 110 × 60 × 60 | 4 |
TI – 90 I S/G | 600 | 125 × 70 × 70 | 4 |
TI – 90 J S/G | 800 | 125 × 80 × 80 | 5 |
TI – 90 K S/G | 1000 | 157 × 80 × 80 | 5 |
TI – 90 L S/G | 1500 (double door) | 146 × 150 × 69 | 10 |
TI – 90 M S/G | 2000 (double door) | 146 × 150 × 92 | 10 |
Optional Features
- CFR 21 part 11 compliance software for data logger
- Data logger
- Data scanner complete with sensor
- Safety digital temperature controller
- Timer
- Motor Speed Controller